Recall Z-0931-2024 ยท KPS System, Tomography, Computed, Emission
Philips North America
BrightView XCT, gamma camera for SPECT; Model Nos.: 6-digit format 882482; 12-digit format 453560462131 453560749161
| Date | Feb 2, 2024 (initiated Dec 15, 2023) |
| Status | Open, Classified |
| Reason | Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation. |
| Root cause | Component design/selection |
| Action | On December 15, 2023, the firm notified affected customers via URGENT Medical Device Correction letters. Customers were instructed to not position a patient's lower limbs directly under the detector below the Center of Gantry bore. Customers should circulate the notice to all users of the device for awareness and retain the letter with affected systems until the correction is complete. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit the customer site and correct the system if necessary. If you need any further information or support con |
| Quantity | 106 US; 241 OUS |
| Distribution | Domestic distribution nationwide. International distribution worldwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K080927 | BRIGHTVIEW VCT IMAGING SYSTEM Philips Medical Systems (Cleveland), Inc., Apr 11, 2008 |