ClearanceBrief
Recall Z-0930-2024 ยท KPS System, Tomography, Computed, Emission

Philips North America

BrightView X, gamma camera for SPECT; Model Nos.: 6-digit format 882478; 12-digit format 453560824741 453560829261;

DateFeb 2, 2024 (initiated Dec 15, 2023)
StatusOpen, Classified
ReasonDetector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
Root causeComponent design/selection
ActionOn December 15, 2023, the firm notified affected customers via URGENT Medical Device Correction letters. Customers were instructed to not position a patient's lower limbs directly under the detector below the Center of Gantry bore. Customers should circulate the notice to all users of the device for awareness and retain the letter with affected systems until the correction is complete. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit the customer site and correct the system if necessary. If you need any further information or support con
Quantity56 US; 46 OUS
DistributionDomestic distribution nationwide. International distribution worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080927BRIGHTVIEW VCT IMAGING SYSTEM
Philips Medical Systems (Cleveland), Inc., Apr 11, 2008