Recall Z-0932-2024 ยท HRS Plate, Fixation, Bone

TriMed Inc.

Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W, VBEAR-3-7N, VBEAR-5-7-N, VBEAR-3-7S, VBEAR-5-7S, VBEAR-7-7S, VBEAR-3-7W, VBEAR-5-7W. Intended to be used as an aid to the treatment of certain types of fractures, non-unions or osteotomies.

DateJan 31, 2024 (initiated Nov 28, 2023)
StatusOpen, Classified
ReasonPlates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes.
Root causeLabeling design
ActionOn 11/28/23, recall notices were emailed to customers who were asked to do the following: 1) Stop further use of the device and return any unused products to the recalling firm. 2) Provide a copy of this recall notice to the hospital/surgeon/physician that is using the device. 3) Complete and return the response form to quality@trimedortho.com Contact quality@trimedortho.com for any additional questions.
Quantity270
DistributionUS: FL, MI, AL, NY, CA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K222637TriMed Wrist Fixation System 3
TriMed, Inc., Dec 21, 2022