Recall Z-1010-2024 ยท KRA Catheter, Continuous Flush

Angiodynamics, Inc.

PulseSpray Infusion System devices including: PULSE SPRAY 5F/135/20, UPN: H787125004045, REF: 12500404; Pulse Spray 3Fx135x20, UPN: H787125004155, REF: 12500415; and PULSE SPRAY 5F/45/10, UPN: H787125004235, REF: 12500423.

DateFeb 7, 2024 (initiated Dec 18, 2023)
StatusOpen, Classified
ReasonRotating adaptor O-ring manufacturing defect impacting seal integrity which may lead to a leak during infusion of medication.
Root causeProcess control
ActionAn Urgent Medical Voluntary Medical Device notices dated December 18, 2023 was issued to distributors and end users with instructions to immediately stop using affected product, remove said product from inventory and return to AngioDynamics, Inc. The notice should be forwarded to all site affected product may have been distributed to and recipients are asked to complete and return the Reply Verification Tracking Form. If affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday - Friday: Eastern Standa
Quantity7 eaches
DistributionUnited States (MD, LA & UT), Austria, Canada and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163356Pulse* Spray Infusion System, Uni*Fuse Infusion System
AngioDynamics, Inc., May 30, 2017