Recall Z-0991-2024 ยท BSZ Gas-Machine, Anesthesia

Getinge Usa Sales Inc

Flow-e Anesthesia System, model 6887900

DateFeb 6, 2024 (initiated Dec 11, 2023)
StatusOpen, Classified
ReasonPotential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
Root causeSoftware design
ActionAn Urgent Medical Device Correction notice dated December 21, 2023 was issued via FedEx. The letter advises customers to examine inventory immediately to determine if you have any affected product. Directions are outlined for 1) Users with system version 4.8 and above with a Flow Anesthesia System connected to Connected Services directly via the Ethernet port on the Flow Anesthesia System (i.e., without a Getinge Connect module) and for 2) Users with system version 4.7 and below with Flow Anesthesia System connected to Remote Services via ethernet cable. Customers are asked to complete and re
Quantity86 (US)
DistributionWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133958FLOW-I ANESTHESIA SYSTEM C20, C30 & C40
Maquet Critical Care AB, Mar 20, 2014