Recall Z-0998-2024 ยท MQL Media, Reproductive

Fujifilm Irvine Scientific, Inc.

Sperm separation media are used to separate motile sperm from other constituents of semen.

DateFeb 6, 2024 (initiated Nov 2, 2023)
StatusOpen, Classified
ReasonThe lower layer component of a sperm separation Kit does not match the certificate of analysis. The kit should contain lower layer (Lot: 0000020686, Shelf Life: 31 December 2024) but incorrectly may contain lower layer (Lot number 9925811106, Shelf Life: 30 November 2023).
Root causeMixed-up of materials/components
ActionOn 11/2/23, recall notices were distributed to customers who were asked to do the following: 1) Review inventory and if the lower layer is incorrect, per the COA, stopusingtheFISIProduct,return/destroyit,andrequestareplacementorcredit. 2)ThisnoticeneedstobepassedontoallthosewhoneedtobeawarewithinyourorganizationortoanyorganizationwherethepotentiallyaffectedProducthasbeentransferred. 2) Complete and return the return response form via email to fisira@fujifilm.com For customers requiring furtherinformationorsupport, contact firm customer service at 1-800-577-6097 orfisitmrequest@fujifilm.com
Quantity246
DistributionUS: OK, CA, DC, NC, NV, TX, MI, IN, NY, UT, WI, MN, PA, WV, NJ, IL, LA, WA, OR, FL, IA. OUS: GB, NL, AR, MY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991341ISOLATE, ISOLATE CONCENTRATE, ISOLATE STOCK SOLUTION
Irvine Scientific Sales Co., Inc., Jul 8, 1999