Recall Z-0970-2024 ยท HNR Forceps, Ophthalmic

Vortex Surgical Inc.

Vortex Surgical ACTU8 FORCEPS, 25 GA, ADAPTIVE, REF VS0740.25

DateFeb 1, 2024 (initiated Dec 18, 2023)
StatusTerminated
ReasonReason for the voluntary recall is residue was identified on the two forceps tips of a customer in Korea. The residue is associated with one lot of forceps tips which were utilized in two lots of ACTU8 Adaptive Forceps Tips occurring in 0.25% of those lots.
Root causeDevice Design
ActionThe firm issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 12/18/2023 via email. The notice explained the problem with the device, risk to health associated with its use, and requested the return of all devices.
Quantity800 devices
DistributionWorldwide - US Nationwide distribution in the states of OH, WA, MN, MI, IA and the country of Republic of Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.