Recall Z-0986-2024 ยท OCW Endoscopic Tissue Approximation Device

Covidien LP

AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material to soft tissue in minimally invasive and open surgical hernia repair procedures Model ABSTACK30

DateFeb 6, 2024 (initiated Nov 16, 2023)
StatusOpen, Classified
ReasonPackaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility
Root causeProcess control
ActionMedtronic issued Urgent Medical Device Recall Letter and Customer Acknowledgement Form to each consignee via mail beginning 16 November 2023. Letter states reason for recall, health risk and action to take: Identify and quarantine all unused and non-expired affected lot N3J1860Y of the ModelABSTACK30 AbsorbaTack" Absorbable Fixation Device listed above. See attachment A forguidance on identifying potentially affected devices. "Return all unused product from the affected lot in your inventory to Medtronic. Please contactrs.covidienfeedbackcustomerservice@medtronic.com for the Return Good Author
Quantity54 units
DistributionUS Nationwide distribution in the states of CA, CO, FL, IL, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090470AUTOSUTURE PROTACK; TACKER SYSTEM; ABSORBATACK FIXATION DEVICES
Covidien, May 14, 2009