ClearanceBrief
Recall Z-0775-2024 ยท BTI Compressor, Air, Portable

Datex-Ohmeda, Inc.

EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

DateJan 30, 2024 (initiated Dec 29, 2023)
StatusOpen, Classified
ReasonGE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstrm Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root
Root causeComponent design/selection
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to consignees on 12/29/2023. The notice explained the safety issue with the device and requested the following actions be taken: GE HealthCare recommends that the EVair and EVair 03 (Jun-Air) compressors are not used to supply air to ventilators for neonatal and infant patients (0-2 years of age). Since the elevated levels of formaldehyde are observed when the compressors are used at higher room temperatures, GE HealthCare is also lowering the maximum room air temperature for operation of the compressors from 40oC (104oF) to 30oC
Quantity12 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.