Recall Z-0773-2024 ยท PHX Shoulder Prosthesis, Reverse Configuration

Tornier, Inc

Stryker Tornier Perform Reversed Glenoid Press-Fit Short Post Ti6AI4V, REF DWJ001, L: 7mm, Cementless Sterile.

DateJan 19, 2024 (initiated Jan 3, 2024)
StatusOpen, Classified
ReasonThe incorrect device is contained in the labeled package.
Root causePackaging process control
ActionThe recalling firm issued letters dated 1/3/2024 via email to sales representatives and distributors and via FedEx to hospitals on 1/3/2024. The letter provided a product description, product issue, potential risks including a photo of the device that should be in the package and a photo of the device that was in the package. The customer actions needed included: (1) Immediately check the inventory for the affected product, remove it from the point of use, and isolate/quarantine it to prevent accidental use; (2) Sign and return the enclosed Business Reply Form by email to confirm receipt of
Quantity50
DistributionDistribution was made to AR, AZ, CA, CO, CT, NC, OH, TN, and VA. There was no military/government distribution. Foreign distribution was made to Australia, France, and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183696Aequalis PerFORM Reversed Glenoid, Aequalis PerFORM + Reversed Glenoid
Tornier, Inc., Jan 30, 2019