Recall Z-0905-2024 ยท LWJ Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented

Howmedica Osteonics Corp.

P7 32MM 1O DEGREE +4MM INSERT Intended for knee replacement Part Number: 6302-6-307

DateJan 31, 2024 (initiated Dec 1, 2023)
StatusOpen, Classified
ReasonProduct manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Root causeNonconforming Material/Component
ActionStryker issued Urgent Medical Device Recall Letter (Product Field Action #: 3306872)through third party Sedgwick via UPS on 12/1/23. Letter states reason for recall, health risk and action to take: Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the affected product in Part/Lot Number Attachment: PFA 3306872 starting on page 3. 4. Co
DistributionNationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.