ClearanceBrief
Recall Z-0896-2024 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Howmedica Osteonics Corp.

SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3708

DateJan 31, 2024 (initiated Dec 1, 2023)
StatusOpen, Classified
ReasonProduct manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Root causeNonconforming Material/Component
ActionStryker issued Urgent Medical Device Recall Letter (Product Field Action #: 3306872)through third party Sedgwick via UPS on 12/1/23. Letter states reason for recall, health risk and action to take: Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the affected product in Part/Lot Number Attachment: PFA 3306872 starting on page 3. 4. Co
Quantity149 units US; 424 units OUS
DistributionNationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051948AVON PATELLO-FEMORAL JOINT PROSTHESIS
Howmedica Osteonics Corp., Oct 21, 2005
K052917STRYKER COMPARTMENTAL KNEE SYSTEM
Howmedica Osteonics Corp., Dec 27, 2005
K972967OSTEONICS SCORPIO TOTAL KNEE UNIVERSAL DOME PATELLAR COMPONENT;OSTEONICS SCORPIO TOTAL KNEE RECESSED PATELLAR COMPONENT
Osteonics Corp., Jan 16, 1998