Recall Z-0778-2024 ยท GEX Powered Laser Surgical Instrument

Accure Acne, Inc

Accure Laser System, Model: PFMS00004

DateJan 20, 2024 (initiated Dec 6, 2023)
StatusOpen, Classified
ReasonLaser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.
Root causeUse error
ActionOn 12/6/23, "Field Safety Alert(s)" were emailed to customers informing them of the following: 1) Following these steps will minimize the likelihood of Region of Interest (ROI) misalignment: Make sure that during treatment that the handpiece tip movement is minimized and to not lift off the tip until the emission is complete and foot pedal released. Remember this order of operations to reduce the potential of misalignment: "Tip down "Pedal down "Emission completes "Pedal up "Tip up 2) Monitor the patient during treatment for unexpected levels of pain. If higher than usual stop treatment and c
Quantity22
DistributionUS distribution to states of: CO, CA, CT, NY, NJ, MN, LA, OH, MA, TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K222109Accure Laser
Accure Acne, Inc., Nov 17, 2022