Recall Z-0722-2024 ยท MQG Accessory, Assisted Reproduction

Cook Medical Llc

MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

DateJan 12, 2024 (initiated Dec 7, 2023)
StatusOpen, Classified
ReasonThe MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
Root causeDevice Design
ActionThe firm notified affected customers on December 7, 2023 via URGENT Medical Device Correction letters. Per the letter, customers can continue to use their MINC+ device whilst awaiting the defect correction, but Cook advises customers to be vigilant and monitor for device alerts. If the device produces audible and visual alerts to indicate an error with temperature, customers must immediately move any dishes to another incubator. If no other incubators are available, the device can be reset to normal operation by switching off mains power to the device for ten seconds then turning the power ba
Quantity14 US; 101 OUS
DistributionDomestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM