Recall Z-0779-2024 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Medtronic Inc.

Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1

DateJan 22, 2024 (initiated Dec 7, 2023)
StatusOpen, Classified
ReasonDuring internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
Root causeNonconforming Material/Component
ActionMedtronic hand delivered an Urgent Medical Device Recall/Retrieval notice to its sole consignee on 12/07/2023. The notice explained the issue and indicated that the device would be collected and a replacement provided.
Quantity1 device
DistributionUS Nationwide distribution in the state of Florida.
FDA recordOpen on fda.gov

510(k) clearances named in this recall