ClearanceBrief
Recall Z-0639-2024 ยท NKB Thoracolumbosacral Pedicle Screw System

Zimmer Biomet Spine Inc.

Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System

DateJan 4, 2024 (initiated Oct 16, 2023)
StatusOpen, Classified
ReasonSpinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.
Root causeNonconforming Material/Component
ActionOn 10/16/23, recall notices were sent to customers asking them to do the following: 1) Locate, quarantine, and return affected screws to the recalling firm. 2) Complete and return the inventory return certificate form via email to Alex.Hayden@ZimVie.com Recall-related questions or concerns can be directed to please email Alex.Hayden@ZimVie.com.
Quantity3
DistributionUS Nationwide distribution in the states of MN, NE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150896Vitality Spinal Fixation System
Zimmer Spine, Inc., Aug 31, 2015