Recall Z-0693-2024 ยท CBK Ventilator, Continuous, Facility Use

Percussionaire Corporation

Distal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)

DateJan 17, 2024 (initiated Mar 28, 2023)
StatusOpen, Classified
ReasonBreathing circuit reservoir bag may be assembled at the wrong attachment point, which will result in less than optimal therapy.
Root causeProcess control
ActionOn 3/28/23, correction notices were emailed to customers who were asked to remove and switch the reservoir bag to the proper location. On 1/24/24, follow-up correction notices were sent to customers who were asked to do the following: 1) Review your inventory, and if affected devices are identified, ensure the reservoir bag is in the proper location. 2) If affected devices were further distributed, send a copy of this notice to your customers. 3) Complete and return the Effectiveness Check Questionnaire via email to customerservice.us@sentec.com If you have any questions, contact the firm's
Quantity195
DistributionUS: RI, CA, TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K905235VDR SINUSOIDAL PERCUSIONATOR FAMILY/RELATE ACCESS
Percussionaire Corp., May 3, 1991