Recall Z-0642-2024 ยท OLO Orthopedic Stereotaxic Instrument

Globus Medical, Inc.

ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100

DateJan 18, 2024 (initiated Dec 7, 2023)
StatusOpen, Classified
ReasonA calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.
Root causeDevice Design
ActionThe firm notified affected customers through URGENT Medical Device Recall Notificatnoi letters dated December 7, 2023. Customers are instructed to immediately return the Flat Panel Fluoroscopy Registration Fixture. For questions related to this recall, please contact Globus Medical Complaint Handling Unit at recall@globusmedical.com.
Quantity9
DistributionDomestic distribution to the following states: CA, CO, MT, SC, WA. International distribution to Germany and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211616ExcelsiusHub
Globus Medical, Inc., Aug 25, 2021