Recall Z-0702-2024 ยท LYR Helicobacter Pylori

Immunostics, Inc.

Detector H. Pylori, Catalog Number IHP-30

DateJan 10, 2024 (initiated Aug 25, 2023)
StatusOpen, Classified
ReasonLabeling for affected lots incorrectly indicates the expiration date is 10/31/2023. However, the correct expiration date is 10/31/2022. Use of the product after its expiration may cause false positive, false negative, or invalid results.
Root causeIncorrect or no expiration date
ActionOn August 25, 2023, the firm initiated the recall with its distributor, who in turn notified affected customers on September 6, 2023. Customers were instructed to dispose of the product and submit a response form directly to Immunostics, the recalling firm. Customers may contact Immunostics by calling toll free at 1800-722-7505, Monday through Thursday, 8:00 AM to 4:30 PM, Eastern Time and Friday 8:00 AM to 4:00 PM Eastern Time or via email at technical@immunostics.com
Quantity10 kits
DistributionProduct distributed to CA and MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980798BIOSIGN H. PYLORI WB-NEW ONE-STEP ANTI-H. PYLORI ANTIBODY TEST
Princeton BioMeditech Corp., Sep 22, 1998