Recall Z-1140-2024 ยท PCH Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System

BioFire Diagnostics, LLC

FilmArray Gastrointestinal (GI) Panel RFIT-ASY-0116 (30 test kit) RFIT-ASY-0104 (6 test kit) Gastrointestinal microorganism multiplex nucleic acid-based assay. In vitro diagnostic.

DateFeb 15, 2024 (initiated Jan 26, 2024)
StatusOpen, Classified
ReasonDue to potential signals of increased false positive Norovirus results when using the gastrointestinal (GI) panel.
Root causeUnder Investigation by firm
ActionOn 01/262024, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via overnight FedEx to customers to inform them of the increased risk of false positive Norovirus results when the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel. Customers are requested to take the following actions: In this context, we request you to take the following actions. Please: - Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred our product.
Quantity64,594 kits (equals 2,303,760 tests)
DistributionWorldwide - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, US Virgin Island, VT, WA, WI, WV, and W
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140407FILMARRAY GASTROINTESTINAL (GI) PANEL
Biofire Diagnostics, May 2, 2014
K143005FilmArray Gastrointestinal (GI) Panel for use with the FilmArray 2.0
Biofire Diagnostics, LLC, Feb 19, 2015
K160459FilmArray Gastrointestinal Panel (GI) for use with FilmArray Torch
Biofire Diagnostics, LLC, Apr 1, 2016
K230404BIOFIRE FILMARRAY Gastrointestinal (GI) Panel
Biofire Diagnostics, LLC, Mar 16, 2023