Recall Z-0683-2024 ยท DQK Computer, Diagnostic, Programmable

Siemens Medical Solutions USA, Inc

Sensis Vibe Hemo

DateJan 9, 2024 (initiated Nov 29, 2023)
StatusOpen, Classified
ReasonThe possibility of the Sensis documentation functionality application to crash.
Root causeSoftware design
ActionThe recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 11/29/23 by mail to consignees to notify them of this recall event. In the case this failure mode occurs during use, consignees are asked to continue the exam by opening a new study for the same patient without adding any Type 1 reporting events. It is instructed that consignees can compensate this by adding free text comments (procedure notes) only and manually inserting the desired information in those comments. The recalling firm will deploy a software update via Update Instruction AX010/23/S to correct the er
Quantity331 units
DistributionDomestic: CA, CT, & FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223812Sensis Vibe (VD15)
Siemens Medical Solutions USA, Inc., Sep 15, 2023