Recall Z-0754-2024 ยท GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered

Cardinal Health 200, LLC

CardinalHealth Suction Canister Liner, MEDI-VAC CRD, REF 65651-510

DateJan 18, 2024 (initiated Dec 21, 2023)
StatusOpen, Classified
ReasonThe above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.
Root causeProcess control
ActionCardinal Health issued an URGENT MEDICAL DEVICE PRODUCT CORRECTION notice to its consignees on 12/21/2023 via overnight delivery. The notice explained the issue with the product and requested the consignee post a copy of Attachment A in the storeroom and clinical areas which provides the following information: Please take the following actions: "" Inspect liner for misalignment with the outer hard canister during setup. The misalignment can be seen visually during setup as the liner not sitting properly within the canister and can be felt by a springing back of the liner from the outer ca
Quantity3505730 units
DistributionUS, Canada, Australia and New Zealand
FDA recordOpen on fda.gov