Recall Z-0707-2024 ยท DPS Electrocardiograph

Baxter Healthcare Corporation

XScribe CP Cardiac Stress Testing System, version 6, with below product names and codes: a) X-SCRIBE CPI KIT WIRELESS USB 608 MHz, 41000-030-50; b) X-SCRIBE CPI KIT WIRELESS USB 2500 MHz, 41000-030-51; c) X-SCRIBE CPI KIT DIRECT CONNECT USB, 41000-030-52; d) X-SCRIBE CPI KIT QUICK CONNECT USB, 41000-030-53; e) XSCRIBE CP KIT WIRELESS, 41000-030-60; f) XSCRIBE CP KIT WIRED, 41000-030-62; g) X

DateJan 10, 2024 (initiated Dec 8, 2023)
StatusOpen, Classified
ReasonPotential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
Root causeSoftware design
ActionBaxter issued an Urgent Medical Device Correction notice to its consignees on 12/08/2023 via USPS first class mail. The notice explained the problem, hazard involved, requested the following actions be taken until a software update becomes available: - Clinicians should evaluate the potential impact of the SCF as described to determine if it should be enabled or disabled during stress testing and provided the method for disabling the setting. - If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them, infor
Quantity1675 units
DistributionUnited States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.