Recall Z-0638-2024 ยท CBK Ventilator, Continuous, Facility Use
Philips Respironics, Inc.
Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100X19 EU2110X15B FR2100X14B FR2110X14B FX2100X15B GB2110X15B IA2100X15B IA2110X15B IN2100X15B IN2100X19 IN2110X15B IT2100X21B IT2110X21B JP2100X16B JP2110X16B KR2110X15B LA2100X15B LA2110X15B ND2110X15B RDE2100X13B R
| Date | Jan 4, 2024 (initiated Nov 27, 2023) |
| Status | Open, Classified |
| Reason | The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2. |
| Root cause | Device Design |
| Action | On November 27, 2023, the firm sent "URGENT Medical Device Correction" letters to dated November 20, 2023, affected customers along with an attached manual addendum for insertion into current manuals. Customers were instructed to keep a copy of the letter and attached addendum with their Trilogy Evo or Trilogy Evo O2 Ventilator IFUs. Distributors were instructed to notify any customers to whom they have distributed affected devices. If you need any further information or support concerning this issue, please contact your local Philips representative: Call 1-(877)-387-3311 or email at patien |
| Quantity | 100,126 (54,691 US; 45,435 OUS) in total |
| Distribution | Worldwide distribution: US (nationwide)-AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Including PR; and OUS (International) including countries of: AE, AR |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181166 | Trilogy Evo Respironics, Inc., Jul 18, 2019 |