Recall Z-0733-2024 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Greiner Bio-One North America, Inc.

VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate, Item #454297, 13x75 grey cap-black ring, non-ridged.

DateJan 16, 2024 (initiated Dec 12, 2023)
StatusOpen, Classified
ReasonTubes have missing additive/anticoagulant resulting in clotting in most cases. A completely clotted sample would lead to no results being reported, which could lead to a delay in treatment or diagnosis. An erroneous glucose or lactose result from a partially clotted sample could lead to inappropriate treatment.
Root causeSoftware design
ActionDistributors were emailed the notification letter dated 12/11/2023 on 12/12/2023 and the other consignees were sent the letter on 12/13-14/2023 via email or USPS. The distributor letter explained the issue, identified the product, the potential health hazard, informed the consignee they should not use the tubes, and requests use of the product immediately be stopped and the product be isolated in their facility. The consignee should assess the risks and consequences of the use of the defective product in accordance with their procedures and take appropriate action. The enclosed Product Disp
Quantity495,600 tubes
DistributionUS Nationwide distribution in the states of CT, FL, IA, IL, MO, NC, NY, PA, TN, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K971239GREINER VACUETTE BLOOD COLLECTION TUBE
Greiner America, Inc., May 12, 1997