Recall Z-0730-2024 ยท LGX Table, Examination, Medical, Powered

Oakworks Inc

Ultrasound EA (Ergonomic Advantage) table version; V2 (Vascular 2 Section) Echocardiography -An Ultrasound examination table

DateJan 13, 2024 (initiated Dec 5, 2023)
StatusOpen, Classified
ReasonBackrest platform stress fracture - the table is not operational in an inclined position
Root causeDevice Design
ActionOakworks Inc. issued "URGENT: MEDICAL DEVICE RECALL/CORRECTION" letter on 12/5/23 via Certified Mail/Email to Distributors and End Users. Letter states reason for recall, health risk and action to take: 1. Immediately examine your inventory of OAKWORKS Inc. Ultrasound EA Table with Echocardiography, or Electric Access Back Supports [left & right] to identify and locate all affected product(s). 2. For tables in use, they remain fully functional if the backrest platform stress fracture has not occurred. For tables that have the stress fracture or break, only use in a flat position until the b
Quantity28 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov