Recall Z-0993-2024 ยท QDS Mers-Cov And Common Respiratory Pathogens Semi-Quantitative And Quantitative Multiplex Nucleic Acid Detection System

BioFire Diagnostics, LLC

FilmArray Pneumonia Panel Plus (Pneumoplus), REF: RFIT-ASY-0143, CED 2797, IVD, Rx Only (30 test kit)

DateFeb 6, 2024 (initiated Dec 14, 2023)
StatusOpen, Classified
ReasonTheir is a potential for false positive results when using Pneumonia panel.
Root causeUnder Investigation by firm
ActionOn December 14, 2023, Biomerieux, issued a "Urgent Medical Recall" to affected consignees via: FedEx . Biomerieux asked consignees to take the following actions: 1. Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. 2. Use another method to confirm positive results for seasonal Coronavirus from the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus prior to reporting the result (e.g., consider resul
Quantity214,080 pouches
DistributionWorldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Bolivia, Brazil,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181324FilmArray Pneumonia Panel plus
Biofire Diagnostics, LLC, Nov 15, 2018