Recall Z-0645-2024 ยท HET Laparoscope, Gynecologic (And Accessories)

Olympus Corporation of the Americas

OLYMPUS ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (10 mm diameter) Model: LTF-S190-10 Used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities, including female reproductive organs.

DateJan 5, 2024 (initiated Nov 20, 2023)
StatusOpen, Classified
ReasonInstructions for Use-User Deviations related to the reprocessing of the videoscope detailed in the Reprocessing Manual Instructions. The steps are required for the proper reprocessing of these videoscopes
Root causeLabeling Change Control
ActionOlympus Medical Systems Corporation issued Urgent Medical Device Correction Letter on November 20, 2023. Letter states reason for recall, health risk and action to take: Olympus requires you to take the following actions: During an evaluation of user implementation of the Instructions for Use related to the reprocessing of the LTF-S190-5 videoscope, Olympus observed deviations from the following reprocessing steps detailed in the respective Reprocessing Manual Instructions. These steps are required for the proper reprocessing of these videoscopes. Therefore, Olympus is reminding users that f
Quantity271 units US
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111425OLYMPUS LTF-S190-10, OLUMPUS OTV-S190, OLYMPUS CLV-S190
Olympus Medical Systems Corp., Apr 20, 2012