Recall Z-0758-2024 ยท FTD Lamp, Surgical
Getinge Usa Sales Inc
Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR
| Date | Jan 19, 2024 (initiated Nov 8, 2023) |
| Status | Open, Classified |
| Reason | A potential for a light system to fall in the operating room. |
| Root cause | Equipment maintenance |
| Action | An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and |
| Quantity | 236793 units in total |
| Distribution | US Nationwide. Global Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K130513 | VOLISTA 400; VOLISTA 600 Maquet S.A.S., May 8, 2013 |