Recall Z-0617-2024 ยท FEB Accessories, Cleaning, For Endoscope

Steris Corporation

Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-010 ASSY, HOOK-UP, OLYMPUS 2-8-014 HOOKUP, OPTIM LT 2-8-030 ASSY, HOOKUP, PENTAX-0K 2-8-040 ASSY, HKUP STORZ 0K 2-8-070 ASSY, HU VISION SCIENCE 2-8-112 ASSY, HOOKUP OLMPUS 1K 2-8-120 ASSY, HOOKUP FUJI 1K 2-8-210 ASSY, HOOK

DateDec 22, 2023 (initiated Nov 13, 2023)
StatusOpen, Classified
ReasonIdentification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures.
Root causeProcess control
ActionOn November 13, 2023, the firm notified customers via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to inspect their inventory for affected product. You should not attempt to use a hookup that cannot be effectively identified. Steris will provide either a replacement hookup or replacement label for all affected hookups. If you have questions regarding this matter, please contact Regulatory_Compliance@STERIS.com.
Quantity2545 US; 992 OUS
DistributionDomestic distribution nationwide. Foreign distribution to Australia Canada China Hong Kong Italy Malaysia Netherlands New Zealand Singapore Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082988MEDIVATORS ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM, RAPICIDE PA HIGH LEVEL DISINFECTANT
Medivators Reprocessing Systems, Feb 13, 2009
K102996MEDIVATORS ADVANTAGE PLUS ENDOSCOPE REPROCESSOR
Medivators Reprocessing Systems, Jan 13, 2011
K172677Advantage Plus Pass-Thru
Medivators, Inc., Dec 19, 2017