Recall Z-0631-2024 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Covidien

Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD

DateJan 3, 2024 (initiated Nov 16, 2023)
StatusOpen, Classified
ReasonDue to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
Root causeSoftware Manufacturing/Software Deployment
ActionOn November 16 and 20, 2023, Covidien LLC (Medtronic Company) issued a "Urgent: Medical Device Correction" 1. Immediately notify all personnel in all care environments in which the Valleylab" FT10 Energy Platform. 2. Update ValleylabTM FT10 Energy Platform to software version 4.0.4. For additional information see customer communication letter 3. Until the software is updated, the Valleylab" FT10 Energy Platform and LigaSure" devices can continue to be used as instructed in the User Guide and per your facility protocols. 4. Complete the attached Customer Confirmation Form and return it as
Quantity47,901 units
DistributionWorldwide distribution - US: Nationwide and surrounding territories.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191601Valleylab FT10 Electrosurgical Platform
Covidien, Jul 12, 2019