Recall Z-0788-2024 ยท ISW Assembly, Knee/Shank/Ankle/Foot, External

Ossur H / F

Power Knee, REF: PKA10001, and PKA10003; with Power Knee Battery, REF: PKA10002

DateJan 25, 2024 (initiated Dec 26, 2023)
StatusOpen, Classified
ReasonKnee batteries may need replacement with conforming batteries to ensure there isn't a failure to turn on the device, or an unexpected shutdown, which is a momentary loss of power to the knee where the user may experience loss of support that can lead to a fall and potential injury.
Root causeProcess control
ActionOn 12/26/23, recall notices were mailed to customers who were asked to do the following: a. Check the serial number of the Power Knee Batteries to confirm the affected units. b. Replace the affected batteries with the new batteries supplied along with this letter. c. To ensure future traceability, please fit each new battery into the designated knee according to the knee Serial Number indicated on the sticker on the back of the battery. d. PASS THIS NOTICE to those who need to be aware within your organization. If you have further distributed this product, please identify your customers, and
Quantity544 Batteries
DistributionUS Nationwide distribution including in the states of MA, SC, MD, TN, ND, VA, IL, NJ, FL, MI, NY, AL, TX, OK, IN, GA, CA, PA, NM, VT, NV, MS, HI, OH, WI, LA, IA, NE, MN, AR, OR, NC, NH, WV, CO, ID, AZ, MO, CT, WA, KY, UT, KS.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.