Recall Z-0710-2024 ยท CAC Apparatus, Autotransfusion

Maquet Cardiovascular, LLC

Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit

DateJan 12, 2024 (initiated Dec 13, 2023)
StatusOpen, Classified
ReasonExpress chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
Root causeMaterial/Component Contamination
ActionOn December 13, 2023, the firm notified customers through "URGENT MEDICAL DEVICE CORRECTION" letters. Customers were instructed to examine their inventory to determine if they had any affected lots of the Atrium Express Dry Suction Dry Seal Chest Drains. The correction only affects the sterile water syringe; do not use this syringe. If customers do not need visualization of active pneumothorax, they may keep the affected Express chest drain and set up without water. If customers wish to use the Express chest drain as intended, they may replace the affected sterile water syringe with a new
Quantity(7,501 cases/45,006 devices US); no OUS
DistributionWorldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201305Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain, Express Dry
Atrium Medical Corporation, Sep 3, 2021