Recall Z-0624-2024 ยท OHQ Multi-Analyte Controls Unassayed

Randox Laboratories Ltd.

SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349

DateJan 2, 2024 (initiated Nov 6, 2023)
StatusOpen, Classified
ReasonA decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.
Root causeProcess control
ActionRandox issued Urgent Medical Device Correction Letter on 11/6/23 to the US Distributor Randox Laboratories-US, Ltd. , Kearneysville, West Virginia, for issuance to the user level. Letter states reason for recall, health risk and action to take: As the control material is still testing positive for HSV Type 1/2 lgM, there is no risk to health. If the control did test negative for this analyte, the affected run should be discarded, and the samples re-analysed. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA
Quantity1 unit
DistributionNY, WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.