Recall Z-0616-2024 ยท IYE Accelerator, Linear, Medical

Landauer

Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

DateDec 22, 2023 (initiated Nov 15, 2023)
StatusCompleted
ReasonMicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
Root causeOther
ActionLandauer Issued an URGENT - Medical Device Recall notice to its consignees on 11/15/2023 via email. The notice explained the issue, potential risk, and requested return of the device.
Quantity584 units
DistributionUS and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192196Medical Radiation Dosimetry System microSTARii
Landauer, Dec 27, 2019