Recall Z-0603-2024 ยท MYL Assay,Enzyme Linked Immunosorbent,Parvovirus B19 Igg

Zeus Scientific, Inc.

ZEUS ELISA Parvovirus B19 IgM Test System, Product Number SM9Z7701G

DateDec 21, 2023 (initiated Oct 31, 2023)
StatusOpen, Classified
ReasonThe organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
Root causeDevice Design
ActionAn URGENT NOTICE Medical Device Recall was sent to customers on 10/31/23. Actions to be Taken: Please identify and discontinue use of the above product immediately. In addition, quarantine all unused product in preparation for returning it to ZEUS Scientific. If you have used all the above product(s), please indicate as such on the attached Medical Device Recall Reply Form. If you have any product with this lot number, please identify, discontinue use, and quarantine all unused product in preparation for returning it. In addition, please indicate the quantity of kits remaining in inventory o
Quantity293 kits
DistributionUS States: CA, MA, MI, NJ, NY, TN, UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.