Recall Z-0576-2024 ยท JAK System, X-Ray, Tomography, Computed
Philips North America Llc
The device is a whole-body CT x-ray system. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. The system also includes signal analysis and display equipment, patient and equipment supports, components, and accessories. Philips Inclusive CT has a 72cm bore and includes a detector array that provides 50cm s
| Date | Dec 18, 2023 (initiated Mar 29, 2022) |
| Status | Completed |
| Reason | Three software issues affecting incorrect image display, error interpreting patient images due to image reporting, and patient exposure to incorrect image/function during clinical use. |
| Root cause | Software design |
| Action | Customers were sent IMPORTANT PRODUCT NOTICE letters dated 03/29/2022 via USPS. Consignees were asked to identify any affected devices at their facility utilizing the provided instructions, refer to the provided Appendix A for specific details regarding the three issue descriptions and work arounds provided to customers, place the notice with the documentation of the system until all issues have been resolved, and circulate this notice to all users of this device so that they are aware. A Philips Field Service Engineer will contact consignees to schedule a software upgrade. Customers with any |
| Quantity | 330 units |
| Distribution | Domestic: AL, AR, AZ, CA, FL, GA, HI, IL, IN, KY, LA, MD, ME, MI, MS, NC, NJ, NM, NY, OH, OK, PA, RI, TN, TX; Foreign: Albania, Argentina, Australia, Austria, Bolivia, Bosnia-Herzegovina , Cambodia, Canada, Chile, Colombia, Costa Rica, Czech Republic, Ecuador, Egypt, Ethiopia, France, Germany, Greec |