Recall Z-0739-2024 ยท DRS Transducer, Blood-Pressure, Extravascular

Smiths Medical Asd Inc

Medex LOGICAL CATH LAB KIT, List Number M20754

DateJan 17, 2024 (initiated Nov 10, 2023)
StatusOpen, Classified
ReasonThere is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 11/10/23 was sent to customers. Required Actions for Users: 1. Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2. Inform potential users of the product in your organization of this notification and complete and return the attached response form to smithsmedical7166@sedgwick.com within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product and/or it has already been
Quantity31,685 total
DistributionWorldwide distribution - US Nationwide and the countries of Canada, UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K820355DISPOSABLE DIAPHRAGM DOMES
Medex, Inc., Mar 4, 1982