Recall Z-0655-2024 ยท DYB Introducer, Catheter
Abiomed, Inc.
Abiomed Kit, 23Fr Introducer, 11cm, Sterile Part of Pump Set (1000323)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 2000342
| Date | Jan 5, 2024 (initiated Nov 7, 2023) |
| Status | Open, Classified |
| Reason | Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis |
| Root cause | Process control |
| Action | Abiomed issued URGENT MEDICAL DEVICE RECALL (REMOVAL) Letter Products Impacted: Introducer Kit for Impella (sold within Impella Pump sets and Individually Packaged) Axillary Introducer Kits on 11/7/23. Letter states reason for recall, health risk and action to take: IF AN IMPELLA PUMP IS CLINICALLY NEEDED, ACTIONS TO BE TAKEN BY THE USER: 1. Prior to usage, carefully examine the outer pouch of the Introducer Kit for holes following the provided instructions in Attachment 2. 2. For non-trayed introducers (Part Numbers 0052-3006): IF THERE IS ANY SUSPICION OF THIS TYPE OF PACKAGING DAMAGE, REFRA |
| Quantity | 84 units within kits |
| Distribution | Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.