ClearanceBrief
Recall Z-0646-2024 ยท DYB Introducer, Catheter

Abiomed, Inc.

Abiomed Axillary Insertion Introducer -Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3006 Individual, and Part of Pump Set (005062, 0550-0008, 1000100, 0052-0011, 1000220)

DateJan 5, 2024 (initiated Nov 7, 2023)
StatusOpen, Classified
ReasonHole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
Root causeProcess control
ActionAbiomed issued URGENT MEDICAL DEVICE RECALL (REMOVAL) Letter Products Impacted: Introducer Kit for Impella (sold within Impella Pump sets and Individually Packaged) Axillary Introducer Kits on 11/7/23. Letter states reason for recall, health risk and action to take: IF AN IMPELLA PUMP IS CLINICALLY NEEDED, ACTIONS TO BE TAKEN BY THE USER: 1. Prior to usage, carefully examine the outer pouch of the Introducer Kit for holes following the provided instructions in Attachment 2. 2. For non-trayed introducers (Part Numbers 0052-3006): IF THERE IS ANY SUSPICION OF THIS TYPE OF PACKAGING DAMAGE, REFRA
Quantity717 single units; 1906 within sets
DistributionNationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192769Impella CP Introducer
Abiomed, Inc., Nov 18, 2019