Recall Z-0589-2024 ยท MWS Stabilizer, Heart

Maquet Cardiovascular, LLC

Acrobat-i Vacuum Positioner System, Model Number XP-5000Z

DateDec 19, 2023 (initiated Oct 31, 2023)
StatusOpen, Classified
ReasonThe Positioner Arm may not tighten or lock during normal use.
Root causeComponent design/selection
ActionAn URGENT MEDICAL DEVICE Removal notification letter dated 10/31/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have received the Acrobat-i Vacuum Positioner System having one or more of the lot numbers that are affected by this medical device removal. Please examine your inventory immediately to determine if you have any of the Acrobat-i Vacuum Positioner Systems with the product codes/lot numbers listed in this notice and remove these from use. Return any unused/unexpired affected product to Maquet/Getinge. Please contact Maquet/Getinge Custo
Quantity6472 units
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.