Recall Z-0608-2024 ยท JIX Calibrator, Multi-Analyte Mixture

Siemens Healthcare Diagnostics Inc

RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 400-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491448

DateDec 21, 2023 (initiated Oct 26, 2023)
StatusOpen, Classified
ReasonRAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
Root causeUnder Investigation by firm
ActionSiemens Healthineers issued Urgent Medical Device Correction POC 24-004.A.US on November 1, 2023 to US customers. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. If you experience this issue, you may request a no-charge replacement or receive a credit from your local Siemens Healthineers representative. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please re
Quantity6611 cartridges US
DistributionNationwide with Worldwide Distribution to: Albania Algeria Angola Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Brunei Darussalam Bulgaria Burkina Faso Cambodia Canada Chile China Colombia Congo, Republic Costa Rica
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122398RAPIDPOINT 500 SYSTEM MEASUREMENT CARTRIDGE WITH CALIBRATORS
Siemens Healthcare Diagnostics, Inc., Oct 12, 2012