Recall Z-0590-2024 ยท IEO Processor, Tissue, Automated

LEICA BIOSYSTEMS NUSSLOCH GMBH

Leica HistoCore PEGASUS (PROCESSOR, TISSUE, AUTOMATED)

DateDec 20, 2023 (initiated Nov 20, 2023)
StatusOpen, Classified
ReasonThere is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or edited protocols.
Root causeOther
ActionLeica issued an URGENT Medical Device Recall notice to its consignees on 11/20/2023 via email. The notice explained the issue with the device and instructed the consignee to ensure the selection of the correct setting of the carryover value for al created or edited protocols and follow the instructions in the IFU.
Quantity98 devices
DistributionUS, Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.