Recall Z-0532-2024 ยท PHX Shoulder Prosthesis, Reverse Configuration

Encore Medical, LP

RSP Glenoid Baseplate, 30mm P2 Porous Coated REF 508-32-204. DJO Surgical shoulder device.

DateDec 12, 2023 (initiated Sep 6, 2023)
StatusOpen, Classified
ReasonCoating hex did not engage with the Straight Mod Hexdriver or with the Hex Head Screw Driver
Root causeComponent design/selection
ActionOn 9/6/2023, the firm sent an email, "SE190 URGENT PLEASE READ-QUARANTINE NOTICE" stating that "identified a potential issue" with the RSP BASEPLATE, 30MM, W/P2 COATING". This email instructed customers to "Please reply all to this email when you've located the part(s) and have it quarantined. We will provide further instruction as soon as possible. Please do not request an RMA at this time". On 11/10/2023, the firm sent via email an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that the RSP Glenoid Baseplate was found to have an undersized hex, preventing engagement wit
Quantity146 units
DistributionU.S.: AZ, CA, FL, IN, KS, KY, ME, MN, MS, NJ, OH, Puerto Rico, RI, SC, TN, TX, VA, and WA O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112069RSP GLENOID BASEPLATE POROUS COATED
Encore Medical L.P., Nov 15, 2011