Recall Z-0517-2024 ยท MUJ System, Planning, Radiation Therapy Treatment

Bebig Isotopentechnik Gmbh

SagiPlan 2.2, CE0344, Rx Only , Brachyterapy Treatment Planning System

DateDec 8, 2023 (initiated Oct 17, 2023)
StatusOpen, Classified
ReasonDue to software malfunction, numerical values may be rounded resulting in inaccurate measurements.
Root causeSoftware Design Change
ActionOn October 17, 2023, Eckert & Ziegler BEBIG GmbH issued a "Urgent Field Safety Notice" via email. Eckert & Ziegler asked consignees to take the following actions: 1. If the Length Unit is set to millimeters, user must carefully review and, if necessary, adjust the Image Sequence Plane Separation value during import when working with slice thicknesses that have two decimals, as shown in Figure 1. Users must consider that this approach is not intrinsically safe, therefore solution 1 or 2 is highly recommended 2. If you have used mm as length unit and may have used two decimal numbers in mm
Quantity2 software licenses in US, 255 software licenses in OUS
DistributionUS: MD OUS: Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141900SAGIPLAN, SAGIPLAN PROSTATE MODULE ADD ON, SAGIPLAN INCLUDING PROSTATE MODULE
Eckert & Ziegler Bebig GmbH, May 20, 2015