Recall Z-0522-2024 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Inc.

Medtronic SynchroMed, Model A10

DateDec 9, 2023 (initiated Jul 13, 2020)
StatusCompleted
ReasonThe previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.
Root causeSoftware Design Change
ActionMedtronic issued a SAFETY NOTIFICATION to its consignees on 07/13/2020 by letter. The notice explained the problem with the device and requested the software be upgraded to version 1.1.342.
Quantity2543 units
DistributionForeign Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.