ClearanceBrief
Recall Z-0525-2024 ยท OFF Central Venous Catheter Tray

MEDLINE INDUSTRIES, LP - Northfield

STERILE PVP SOLUTION, REF DYNDA1907

DateDec 11, 2023 (initiated Nov 3, 2023)
StatusOpen, Classified
ReasonThe product description on the labeling includes the word "Sterile" or STRL", however the contents of the solution inside the foil sachet is non-sterile.
Root causeUnder Investigation by firm
ActionMedline Industries notified its consignees on 11/03/2023 via by email and US Mail (first class). The notice explained the issue and requested the destruction of any affected inventory. Distributor were directed to notify their customers.
Quantity8250 units
DistributionUS Nationwide distribution in the states of CT, IA, LA, MA, PA, CA, NY, FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.