Recall Z-0550-2024 ยท FKX System, Peritoneal, Automatic Delivery

Baxter Healthcare Corporation

Baxter Amia Automated Peritoneal Dialysis (APD) Cycler Set, REF 5C5479

DateDec 14, 2023 (initiated Nov 15, 2023)
StatusOpen, Classified
ReasonThere is to an increase in complaints related to failure alarms for Wet Cassette Integrity Test occurring on the Amia devices.
Root causeProcess control
ActionBaxter issued an Urgent Medical Device Recall notice to its consignees on 11/15/2023 by US Mail. The notice explained the issue, potential risk, and requested the following: Isolate and cease all use of the affected units pending return.
Quantity339,936 devices
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151525Amia Automated PD System with Sharesource
Deka Research and Development, Oct 9, 2015