ClearanceBrief
Recall Z-0547-2024 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Ortho-Clinical Diagnostics, Inc.

VITROS XT 7600 Integrated System Product Code 6844461 Running Software Version 3.8.0

DateDec 13, 2023 (initiated Oct 10, 2023)
StatusOpen, Classified
ReasonDuring planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the rele
Root causeUnder Investigation by firm
ActionAn URGENT PRODUCT CORRECTION NOTIFICATION letter dated 10/10/23 was sent to customers. REQUIRED ACTIONS " Review the Customer Actions for the Condition Codes described in this notification. " Acknowledge your understanding of this notification by completing the enclosed Confirmation of Receipt form no later than October 18, 2023. " Save this notification with your User Documentation or post this notification by each VITROS 3600/4600/5600/XT 3400/XT 7600 System until this issue is resolved. " If your laboratory has experienced the issue described in this notification and you have not already d
Quantity216 units
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182063VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, VITROS Chemistry Products TBIL Slides, VIT
Ortho-Clinical Diagnostics, Inc., Oct 30, 2018