Recall Z-0306-2024 ยท JAK System, X-Ray, Tomography, Computed

Philips North America

Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

DateNov 15, 2023 (initiated Oct 9, 2023)
StatusOpen, Classified
ReasonWhen preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted.
Root causeUnder Investigation by firm
ActionAn URGENT Medical Device Correction notification letter dated 10/9/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users. Below are short-term precautions to take until the permanent solution is installed. Leave the audible preparation indication alert turned on, set as default in the factory settings, and wait for the audible alert before pressing the foot pedal. If you have disabled the audible alert, you can turn it back on by following instructions provided in Section 4: Preparing for an Examination of the IFU. Wait
Quantity1 unit
DistributionUS Nationwide distribution in the state of MN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212875Spectral CT on Rails
Philips Medical Systems Nederland B.V., May 13, 2022